Circulators offer many application possibilities, e.g. routine laboratory applications,temperature control of samples, incubations, dental applications, material testing, corrosion tests, tempering of cultures as well as temperature testing of food and luxury articles. The ‘TCB” baths can be used as circulator or as a water/oil bath.
Main Characteristics
The set point can be set in steps of 0.1° in the range from 0°C up to 200°C (32…392°F). The accuracy on the display is displayed in 0.1°C. However, the controller has an internal accuracy of 0.01°C. Two decimal readout is also available via the RS232 and the free Tamcom software. Due to the “self heating” of the stirring mechanism and when using a top lid, the minimum temperature lies approximately 5K above the ambient temperature. When using the standard built-in cooling coil the minimum operating temperature lies approximately 5K above the temperature of the cooling liquid. In general the minimum temperature lies around 10°C. Lower temperatures generate problems with built up of condensate in the inner insulation of the bath.