Density is one of the important physical parameters of liquid chemical products. Determination of density can distinguish
liquid substances with similar chemical composition but different densities,
identify the purity of liquid products and the concentration of some solutions.
Therefore, density is one of the quality control indicators of many liquid
products in chemical product inspection.
Liquid density meter is an important tool for measuring liquid
density, which can accurately measure liquid density. It has the advantages of
high measurement accuracy, wide range, simple structure, easy operation and so
on. It is widely used and plays an important role and value in many fields.
Liquid densitometer has such an important role and value. So how to ensure that the density meter value is really accurate and reliable? The only way is to calibrate according to the national metrological technical specification to ensure that the measurement value is accurate and traceable.
|
The DM-01 density measuring standard
device is an automatic calibration instrument with wide measuring range, no
need for liquid, and high safety factor that can be traced to the quality and
length standards. A high precision
thermostatic chamber with temperature stability better than ±0.01oK is placed
on a height adjustable base, and a transparent cylinder containing tridecane is
placed in a high precision thermostatic chamber. During the verification process,
the densitometer (glass float meter) is suspended on the balance through a fine
wire, and the densitometer scale is partially immersed in tridecane. Adjust the
height of the base, change the scale line immersed in tridecane, and with the
balance weighing value, tridecane density value, external weight (or stainless
steel ring) and other parameters, accurately calculate the corresponding
density value of the scale line. In theory, by tracing the balance, tridecane
density, external weight (or stainless steel ring) quality, the device can be
traced. In practice, the device is equipped with a quartz ring as a standard,
and the quartz ring is sent to the Chinese Academy of Metrology regularly to
complete the traceability. The instrument uses the image sensor to quickly and
accurately mark the line of the ruler, uses the PLC to control the lifting of
the base, collects the weighing data of the balance, calculates the density
value, and can automatically generate the verification report.
Applicable
It is a standard device for the
traceable verification of density meters by metrological technical
institutions. It completely solves the problem that the traditional density
measurement method uses toxic and harmful substances to mix liquid, uses naked
eye observation and consumes time.
The device has been successfully
applied to calibration institutions such as Guangxi and Jinan.
|