Description
VAC
capillary viscometer automatic calibration device, based on computer image
processing algorithm, using automation and intelligent application technology,
to achieve automatic calibration of capillary viscometer. The
calibrated viscometer includes four categories: Ubbelohde Viscometer,Pinkevitch Viscometer, Cannon-Fenske Routine Viscometer ,Cannon-Fenske Opaque Viscometer. Calibration, each station can be placed in different
types, different models of capillary viscometers, calibration process is
independent of each other, can be asynchronous calibration, do not affect each
other.
Advantages
-
High efficiency: each station asynchronous calibration, do not
affect each other
-
Flexibility: different types of viscometers can be placed at
each station
- Practicability: All four categories of viscometers specified in
the regulations can be checked
-
Reliability: Unique image processing algorithm, accurate and
reliable identification timing
-
Economy: The user can choose 1~6 stations, and the subsequent
stations can be added
-
High return: The viscosity project enhances user viscosity, and
the output value is high
Specifications
Working station number
|
1~6 Optional typical value 6, with the precision transparent constant temperature bath bath opening and window size
|
Calibration method
|
Automatic identification and verification of image sensor
|
Viscometer category
|
Wood type, flat type, Fin type, countercurrent type
|
Time repeatability
|
Electronic timing, according to JJG 155-2016 Table 1 requirements
|
Constant temperature range
|
20~180°C
|
Temperature fluctuation
|
No more than ±0.01°C/h
|
Temperature field uniformity
|
No more than ±0.02°C
|
Thermometer resolution
|
Better than 0.001°C
|
Calculate the time degree value
|
Electronic timing, better than 0.01s, the daily difference does not exceed ±0.05s
|
Degree of automation
|
Automatic suction, automatic recognition, automatic timing function
|
Measurement and control software function
|
Parameter setting, process recording, image display, generating report, etc. Voltage 220VAC, 50 Hz
|
|
VAC-A1
|
Quick adjust vertical mechanism, optional accessories
|
VAC-A2
|
Automatic filling mechanism, optional accessories
|
|
Applications
Verification and calibration of capillary viscometer
Capillary viscometer during verification
Automatic determination of oil motion viscometer
Environmental parameters
Ambient temperature range: 0 to 40°C
Ambient temperature stability: ±2°C/d
Relative humidity: 0~ 80%RH, no condensation
|
|
|